Recall Z-1319-06 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer America Inc

D711, Reference Membranes, Part #942-058, used with the ABL835 Blood Gas Analyzers.

DateAug 2, 2006 (initiated Jun 17, 2005)
StatusTerminated
ReasonThe recalled membranes cause the status value of the pH and electrolyte electrodes to be offset causing readings outside of the acceptable range.
Root causeOther
ActionThe recalling firm notified their distributor/customers of this recall, via telephone on 6/17/2005, at which time, the customers were asked to destroy any and all suspect reference membranes found on-site and at their subaccount customer locations.
Quantity13 boxes/4 membranes per box
DistributionNationwide, including TX, OK, MI, CO, NY, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041874ABL800 FLEX
Radiometer Medical Aps, Oct 8, 2004