Recall Z-1182-06 ยท IWB System, Radiation Therapy, Radionuclide

MDS Canada Inc. DBA MDS Nordion

Theratron Phoenix Cobalt Radiotherapy

DateJun 27, 2006 (initiated Jul 8, 2004)
StatusTerminated
ReasonThe source locking device may not fit properly through the smaller opening in the new field light shield which could prevent the source from being returned manually under emergency conditions.
Root causeOther
ActionA User Bulletin (CUB-2004-01dated 7/5/04) was issued to affected users on 8/24/04 and 8/31/04. The bulletin provided 2 options for corrective action.
Quantity4 units
DistributionThe products were shipped to medical facilities in FL and KS. The product was also shipped to Kuwait, Spain, Honduras, Romania, France, Grrece, Bangladesh, Germany, Nepal, Canada, England, Thailand, Phillippines, Mexico, Columbia, India, Switzerland, Syria, Trinidad & Tobago, Ireland, Turkey, Algeri
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K863180THERATRON-PHOENIX
Aecl Medical, Oct 27, 1986