ClearanceBrief
Recall Z-1370-06 ยท LFR Glucose Dehydrogenase, Glucose

Roche Diagnostics Corp.

ACCU-CHEK Voicemate System, Advantage Blood Glucose Meter and Voice Unit - English language version for the blind or visually impaired, Cat. no 2030802. Ref. no. 12030802001 with meter model no. 768.

DateAug 5, 2006 (initiated Jun 22, 2006)
StatusTerminated
ReasonThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
Root causeOther
ActionA recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.
Quantity55,189 meters (for all Recalls Z-1354-1371-06)
DistributionWorldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982079ACCU-CHECK VOICEMATE
Boehringer Mannheim Corp., Aug 28, 1998