Recall Z-1314-06 ยท CGZ Electrode, Ion-Specific, Chloride

Radiometer America Inc

D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer.

DateAug 1, 2006 (initiated Feb 1, 2005)
StatusTerminated
ReasonThe CI membrane units generate too low values for status and sensitivity during calibration.
Root causeOther
ActionThe recalling firm sent a letter, dated 2/05, to all of their consignees.The letter instructed the customers to discontinue use of the product and discard any remaining stock which they had on hand. The letter was accompanied by replacement chloride membranes to replace the customer''s portion of the suspect lot.
Quantity85/4 unit boxes
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980130ABL700 SERIES BLOOD GAS , CO-OXIMETRY, ELECTROLYTE & METABOLITE SYSTEM
Radiometer America, Inc., Feb 9, 1998