Recall Z-1368-06 ยท CGA Glucose Oxidase, Glucose

Roche Diagnostics Corp.

GTS Advantage Meter. Cat. no. 424. Ref. no. 03144780001 with meter model no. 777.

DateAug 5, 2006 (initiated Jun 22, 2006)
StatusTerminated
ReasonThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
Root causeOther
ActionA recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.
Quantity55,189 meters (for all Recalls Z-1354-1371-06)
DistributionWorldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.