Recall Z-1367-06 ยท CGA Glucose Oxidase, Glucose

Roche Diagnostics Corp.

AccuData GTS Plus AccuData GTS For Blood Glucose Monitoring and Automated Data Management, contents Accu-Chek Advantage repalcement Meter and module, cat. no. 454; Ref. no. 03144798001 contains meter model no. 777.

DateAug 5, 2006 (initiated Jun 22, 2006)
StatusTerminated
ReasonThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
Root causeOther
ActionA recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.
Quantity55,189 meters (for all Recalls Z-1354-1371-06)
DistributionWorldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K924475ACCUDATA GLUCOSE TEST STATION
Boehringer Mannheim Corp., Oct 13, 1992