Recall Z-1366-06 ยท CGA Glucose Oxidase, Glucose

Roche Diagnostics Corp.

AccuData GTS Plus for Blood Glucose Monitoring and Automated Data Management including GTS Plus with Accu-Chek Advantage blood glucose meter, Cat no. 3000249 and/or GTS with Accu-Chek Advantage blood glucose meter, Cat. no. 404, and/or Replacement GTS with Accu-Chek Advantage blood gulcose meter, Cat. no. 404R, and/or Replacement Base Unit with Advantage Module, Cat. no. 3426050. Ref. nos. 030002

DateAug 5, 2006 (initiated Jun 22, 2006)
StatusTerminated
ReasonThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
Root causeOther
ActionA recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.
Quantity55,189 meters (for all Recalls Z-1354-1371-06)
DistributionWorldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K924475ACCUDATA GLUCOSE TEST STATION
Boehringer Mannheim Corp., Oct 13, 1992