Recall Z-1365-06 ยท LFR Glucose Dehydrogenase, Glucose

Roche Diagnostics Corp.

ACCU-CHEK HQ for Blood Glucose Monitoring and Automated Data Management and Communication; Contents include two Accu-Chek Advantage monitors including Cat. no. 2138018 and Cat. no. 2138026. Ref. nos. 12138018001 and 12218542001 both include include meters with model number 777.

DateAug 5, 2006 (initiated Jun 22, 2006)
StatusTerminated
ReasonThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
Root causeOther
ActionA recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.
Quantity55,189 meters (for all Recalls Z-1354-1371-06)
DistributionWorldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983047ACCU-CHEK HQ SYSTEM
Boehringer Mannheim Corp., Feb 5, 1999