ClearanceBrief
Recall Z-1259-06 ยท LXG Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use

Beckman Coulter Inc

Cytomics FC 500 Flow Cytometry System Part Numbers 626553 (TN; FC500 with UPS), 6605627 (TN; 5 CLR, FC500 (100V)), 6605628 (TN; 5 CLR, FC500 (120V)) , 6605629 (TN; 5 CLR, FC500 (220V)), 6605630 (TN; 5 CLR, FC500 (240V)) with CXP Software Versions 2.0 and 2.1

DateJul 26, 2006 (initiated May 22, 2006)
StatusTerminated
ReasonBeckman Coulter has determined that under certain conditions a system error generated during a Pause and Rotate routine may result in a sample misidentification with a risk of reporting erroneous results.
Root causeOther
ActionA Product Corrective Action letter was sent via US mail the week of May 22, 2006 to all FC500 customers. They were informed that under certain conditions a system error generated during a Pause and Rotate routine may result in a sample misidentification with a risk of reporting erroneous results. The letter includes the sequence in which the issue occurrs: 1) During the acquistion of a tube, the Pause and Rotate Button is selected. 2) A system error is generated during the Pause and Rotate routine (The error message table may be viewed by going to the CXP Help Menu, select the CXP Help Menu
Quantity424
DistributionNationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.