Recall Z-0001-2007 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE OEC Medical Systems, Inc

GE 4 inch Anterior Cervical Post, GE Part Number 1006385 or 1006385-NAV, GE Healthcare OEC Medical Systems (an accessory used with the InstaTrak 3500 plus and 9800 C-Arm Navigation systems)

DateOct 3, 2006 (initiated May 31, 2006)
StatusTerminated
ReasonWeld defect- the weld between the post body and the screw was not properly formed and can fail, even when minimal force is used.
Root causeProcess control
ActionConsignees were notified by letter on 05/31/2006
Quantity143 units
DistributionNationwide, including one VA facility in PA. No military distribution. Foreign distribution to Canada, China, Italy, Japan, Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K024012OEC 9800 E/CV+ DIGITAL MOBILE IMAGING SYSTEM
Ge Oec Medical Systems, Mar 20, 2003