Recall Z-1246-06 ยท GJS Test, Time, Prothrombin

Instrumentation Laboratory Co.

HemosIL RecombiPlasTin. Prothrombin Time Test. Each RecombiPlasTin kit consists of: RecombiPlasTin (RTF): 5 x 8 mL or 20 mL vials of lyophilized recombinant human tissue factor, synthetic phospholipids with stabilizers, preservative and buffer; and RecombiPlasTin Diluent (RTF Diluent): 5 x 8 or 20 mL vials of an aqueous solution of calcium chloride, polybrene and a preservative. --- FOR IN VITRO

DateJul 18, 2006 (initiated Mar 7, 2006)
StatusTerminated
ReasonFirm received 2 complaints concerning incorrect prothrombin time (PT) results while using the reagents on samples of patients taking the antibiotic CUBICIN (Daptomycin for injection).
Root causeOther
Action"Urgent Product Notifications" were sent to IL customers on 3/07/06 through its distributor, Beckman Coulter.
Quantity8 mL: 10,473; 20 mL: 5110 (all Domestic)
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043184HEMOSIL RECOMBIPLASTIN
Instrumentation Laboratory CO, Dec 22, 2004