Recall Z-1244-06 ยท MOI System, Test, Immunological, Antigen, Tumor

Abbott Laboratories MPG

CA 15-3 Controls, list 9C08-10; an In Vitro diagnostic for use with the AxSYM or IMx CA 15-3 reagents, calibrators and instruments; each kit contains 2 bottles with 115D8:DF3 reactive determinants (human) prepared in TRIS buffer with protein (bovine) stabilizers to yield the following assay values (U/ml): Control L - 35 and Control H - 150; Abbott Laboratories, Abbott Park, IL 60064

DateJul 18, 2006 (initiated Jun 9, 2006)
StatusTerminated
ReasonThe CA 15-3 Controls replicates out of range high resulting in invalid runs.
Root causeOther
ActionAbbott initiated a field correction recall by letter dated 6/9/06 and sent on 6/12/06 to all customers receiving the affected controls. The accounts were informed of the increase in invalid runs due to controls outside of the package insert range with this lot of CA 15-3 Controls. The accounts were requested to assess the inventory for the affected lot of controls, discontinue use of the lot and destroy the remaining inventory of the lot, recording the number of kits on the customer reply form and faxing it to Abbott by 6/21/06, indicating that the letter was received and understood. Credit w
Quantity1794 kits
DistributionWorldwide, including U.S.A., Puerto Rico, Canada, Mexico, Ecuador, Chile, Argentina, Brazil, Germany, Japan, SIngapore, Hong Kong, Thailand, South Korea and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K963926AXSYM CA 15-3
Abbott Laboratories, Sep 15, 1997
K964407IMX CA 15-3
Abbott Laboratories, Nov 10, 1997