Recall Z-1191-06 ยท CBK Ventilator, Continuous, Facility Use

Respironics, Inc.

NeoPAP Neonatal CPAP/Humidification System (Ventilator, Continuous). Model number 1025310

DateJul 1, 2006 (initiated Sep 19, 2005)
StatusTerminated
ReasonGas input pressures over 62 psi may cause CPAP pressure oscillation
Root causeOther
ActionThe recalling firm issued a recall letter to their customers dated 9/19/05 informing them of the problem and the corrective action. (fitting unit with external 50 psi pressure regulator for both air and oxygen)
Quantity232 units
DistributionWorld wide (including medical facilities in AZ, CA, CO, GA, IA, IL, MA, MD, MN, MO, NC, NE, NH, NJ, OH, OK, OR, PA, PR, TN, TX, VA, WI, and WV. Japan, Ireland, Great Britain, Sweden, Portugal, France, Chile, Argentina, Hong Kong, South Africa, Greece, Egypt, Saudia Arabia, Norway, Spain, Belgium an
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040862GUARDIAN NEONATE CPAP/HUMIDIFICATION SYSTEM
Caradyne, Ltd., Jun 30, 2004