ClearanceBrief
Recall Z-1250-06 ยท CBK Ventilator, Continuous, Facility Use

Hamilton Medical Inc

Hamilton brand RAPHAEL Silver Ventilator (Software Version 2.2xS) Product is distributed by Hamilton Medical, Inc., 4990 Energy Way, Reno, NV 89502

DateJul 22, 2006 (initiated May 24, 2006)
StatusTerminated
ReasonALARM FAILURE-Following an oxygen cell calibration, the user may inadvertently and unknowingly disable the alarm system.
Root causeOther
ActionOn May 24, 2006 the manufacturer initiated this recall. The Recalling establishment/US Initial Importer/Distributor-received the manufacturer''s notice on May 31, 2006 and initiated their recall on June 1, 2006. The US distributor has issued a Medical Device Field Correction.
Quantity2,385 for all models
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022679RAPHAEL
Hamilton Medical AG, Oct 22, 2003
K052863RAPHAEL COLOR
Hamilton Medical AG, Jan 6, 2006