Recall Z-1242-06 ยท MKJ Automated External Defibrillators (Non-Wearable)

Kimberly-Clark Corporation

Kimberly-Clark/Ballard medical R2 triangle shaped non-radiolucent multifunction electrode products for adults.

DateJul 28, 2006 (initiated May 18, 2006)
StatusTerminated
ReasonThe tinfoil in the triangle electrodes has the potential to develop cracks which may hinder electrical flow and product performance during defibrillation.
Root causeDevice Design
ActionConsignees were first contacted by phone with follow up letter sent via fax or mail on May 18, 2006. International consignees will be notified through the firm's regional offices.
Quantity95,639 electrode sets
DistributionWorldwide ( including USA, Switzerland, Germany, UK, Canada, Australia, Netherlands, Italy, Greece, Spain, Hong Kong, Hungary, Denmark, Singapore, New Zealand, Ireland, Korea)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K935371R2 PADS, R2 CATH PADS, R2 PACE PADS, R2 DISPERSIVE PAD