Recall Z-1100-06 ยท JHI Visual, Pregnancy Hcg, Prescription Use

Applied Biotech Inc

Signify hCG Serum/Urine 30 Test Kit, Abbott List No. 07K09-30, ABI List No. 6026KSI, Lot# 84936, Exp 02/10/2007

DateJun 9, 2006 (initiated May 5, 2006)
StatusTerminated
ReasonCertain kits from the product lot may contain test devices which contain incorrect reaction strips. Use of such devices may produce false negative results with patient samples at and above the limit of detection of the product.
Root causeOther
ActionDistribution records indicate that the above product lot was produced for sale to Abbott Laboratories and Abbott Laboratories received all of the product lot. A recall certified recall letter was mailed 05/08/06. Abbott was also instructed if they have further distributed any of the referenced product lot, please see that their accounts are contacted, advise them of the urgent notification, and instruct them to destroy any of the affected product lot.
Quantity1198 Kits (35,940 Test)
DistributionIn US : IL
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.