Recall Z-1237-06 ยท MIH System, Endovascular Graft, Aortic Aneurysm Treatment

W L Gore & Assoc, Inc.

GORE TAG THORACIC ENDOPROSTHESIS System, Endovascular Graft, Aortic Aneurysm Treatment, Catalogue Number TG3415, 34mm x 15cm

DateJul 13, 2006 (initiated Apr 28, 2006)
StatusTerminated
ReasonTwo lots of uniquely different product were switched and mislabeled with the wrong length. 34mm x 10cm devices were mislabeled as 34mm x 15cm and vice versa.
Root causeOther
ActionAll the letters were sent via Federal Express to each Gore Field Sales Associate (FSA) to be hand-delivered to the Physician and to the hospital Risk Manager. A cover note to the FSA directing them regarding the hand delivery of the letter. Also included is a memo to the FSA with a document that the FSA is directed to complete, sign/date and return to Gore documenting hand-delivery of each of these letters. These letters were to be hand delivered on 18 May, or 19 May, 2006.
Quantity10
DistributionNationwide to: AL, CT, KS, MA, & MN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.