ClearanceBrief
Recall Z-1454-06 ยท CCH

Pointe Scientific, Inc.

Liquid AutoHDL Cholesterol Reagent Set, Manufactured for Pointe Scientific, Inc.; Catalog Nos. HH945-240, HH945-480, H7545-40, H7545-80, H7545-320 and H7545-1000. A homogeneous method for the direct determination of high density lipoprotein (HDL) cholesterol in serum or plasma.

DateAug 30, 2006 (initiated Jun 12, 2006)
StatusTerminated
ReasonUnexpected changes in QC and proficiency results due to QC and proficiency material matrix.
Root causeOther
ActionConsignees were notified via recall letter dated 6/12/06 to cease using the product, conduct a recall to the user level, destroy product on hand and to notify the firm via the response form of the amount destroyed for replacement purposes. International accounts were notified via email.
QuantityTo be determined
DistributionWorldwide-USA and Greece, India, Korea, Lebanon, Pakistan, Poland, Singapore, Trinidad, Turkey, United Arab Emirates and Venezuela.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K992002AUTOHDL CHOLESTEROL REAGENT SET
Pointe Scientific, Inc.,, Aug 5, 1999