ClearanceBrief
Recall Z-1174-06 ยท MYN Analyzer, Medical Image

Riverain Medical Group

RapidScreen RS-Digital, Medical Image Analyzer.

DateJun 27, 2006 (initiated May 4, 2006)
StatusTerminated
ReasonThe device has an actual sensitivity of 58.2% and a specificity of 3.9% (false positives per image). The product is approved to have a sensitivity of 63.3% and a specificity of 5.0. Thus, there is a difference of -5.1% in sensitivity and +1.1 in specificity.
Root causeOther
ActionThe recalling firm sent a letter dated 5/4/06 to medical facilities, sent a letter dated 5/5/06 to distributors, and sent a letter dated 5/10/06 to GE Healthcare, Waukesha, WI.
Quantity72 devices
DistributionNationwide. The product was distributed to consignees located in IL, CA, MO, CO, WA, MT, ID, OR, NC, TX, TN, FL, UT, MN, OK, WI, OH, AZ, IA, and PA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.