Recall Z-2329-2008 ยท JQP Calculator/Data Processing Module, For Clinical Use

SCC Soft Computer

SoftPath GUI Release Software Versions 3.17 and 4.1. Used in pathology laboratory for receiving accessing specimen reports, billing and management reports for surgical, GYN, NON-GYN and Autopsy cases.

DateSep 22, 2008 (initiated Feb 23, 2006)
StatusTerminated
ReasonLoss of text misrepresented individual tissue diagnosis on the final diagnosis print out from the HIS system software. The specimen heading "J" was removed and the diagnosis appeared under the specimen heading "I" as a result the specimen with the heading "I" had two diagnosis.
Root causeOther
ActionSoftware corrections were sent as of 10/30/2003 for clients to upgrade to, the correction software patch is in versions 3.17.5.16, 3.17.6.10, 3.17.7.3 and 4.1.1.0. The firm did not inform the non-affected consignees of the failure because the firm concluded that the associated risk was low.
Quantity335
DistributionWorldwide Distribution including USA and the country of Canada.