Recall Z-1190-06 ยท CAD Vaporizer, Anesthesia, Non-Heated

Southmedic Inc

Vapofils with Stainless Steel Block Ends, Item Number V0507S (Sevofluane). The vapofil is a keyed anesthetic transfer tube that transfers the anesthetic from the bottle to the vaporizer.

DateJul 4, 2006 (initiated Apr 21, 2006)
StatusTerminated
ReasonSouthmedic has initiated this recall because their Japanese distributor received field reports of leaking vapofils. Leakage may cause non-toxic spills in the hospital environment, distraction of practitioners during conduct of anesthesia and surgery, and may damage certain plastic equipment. Leakage may also expose patients to low concentrations of volatile anesthetics.
Root causeOther
ActionRecall letters were sent to both US consignees via US Postal Service delivery with signature receipt required on April 21, 2006. Each letter included a Recall and Effectiveness Check Form for each consignee to fill out and fax to Southmedic. U.S. and international consignees were instructed to examine inventory immediately to identify if they have any product on hand, discontinue the distribution of these products and return all affected products to Southmedic. Consignees were also instructed to immediately notify their customers if they have further distributed these products.
Quantity186 in US (403 Worldwide)
DistributionUS in FL & NJ Worldwide in Australia, Belgium, Ecuador, Germany, Japan, Netherlands, Sri Lanka, Taiwan, & United Arab Emerates
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K945993VAPOFIL