ClearanceBrief
Recall Z-1097-06 ยท FZP Clip, Implantable

Teleflex Medical

Weck Hem-o-lok L Polymer Ligating Clips; Rx, sterile, large size, non-absorbable polymer ligation clips; 6 clips per cartridge, 14 cartridges per sales unit, 12 sales units per case; Mfg. by Weck Closure Systems, Research Triangle Park, NC 27709; REF 544240

DateJun 9, 2006 (initiated Apr 12, 2006)
StatusTerminated
ReasonThe Hem-o-lok ligating clips may become dislodged following ligation of the renal artery after laparoscopic donor nephrectomy. Hem-o-lok ligating clips are now contraindicated for use in ligating the renal artery during laparascopic nephrectomies in living donor patients.
Root causeOther
ActionTeleflex Medical sent Important Product Safety Information letters dated 4/18/06 to all of their customers who have purchased Hem-o-lok Ligating Clips from Teleflex. The letters informed the healthcare providers that Teleflex has been made aware of incidents in which the ligating clips (sizes L and XL) were reported to have become dislodged following ligation of the renal artery after laparoscopic donor nephrectomy. Although none of the incidents appear to have involved any defect in or malfunction of the ligating clips, misapplication of the Hem-o-lok clips during such laparascopic procedur
Quantity5,573,400 clips
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003337HEM-O-LOK SMX, MODEL 544220; HEM-O-LOK ML, MODEL 544230; HEM-O-LOK MLX, MODEL 544240
Weck Closure Systems, Dec 21, 2000