ClearanceBrief
Recall Z-0991-06 ยท DQY Catheter, Percutaneous

Medtronic, Inc.

CATH. GUIDE 8F SR4.0 W/SH LAUN CATHETER LA8SR40SH LA 8F 100CM SR40 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peripheral vascular system. Product Code: LA8SR40SH

DateJun 1, 2006 (initiated Apr 5, 2006)
StatusTerminated
ReasonPotential for non-sterility due to loss of package integrity
Root causeOther
ActionMedtronic notified accounts and sales representatives on 4/05/06 advising the recalled product be segregated and held for return to the sales rep. and documented on the reconciliation form.
Quantity56 units
DistributionNationwide Foreign: Austria, Australia, Bahrain, Bulgaria, Canada, Czech Republic, Denmark, Germany, Hungary, Italy, Jordan, France, Greece, Japan, Hong Kong, India, Italy, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0234028F LAUNCHER GUIDE CATHETER
Medtronic Vascular, Nov 6, 2002