ClearanceBrief
Recall Z-1356-06 ยท CGA Glucose Oxidase, Glucose

Roche Diagnostics Corp.

ACCU-CHEK Advantage Blood Glucose Meter, Cat. No. 2137500; Ref. no. 0385633001 with meter model no. 768.

DateAug 5, 2006 (initiated Jun 22, 2006)
StatusTerminated
ReasonThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a strip error.
Root causeOther
ActionA recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.
Quantity55,189 meters (for all Recalls Z-1354-1371-06)
DistributionWorldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K930979ACCU-CHEK ADVANTAGE BLOOD GLUCOSE SYSTEM
Boehringer Mannheim Corp., Sep 8, 1993