ClearanceBrief
Recall Z-1084-06 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

Guidant Endovascular Solutions, Inc.

ABSOLUTE .035 Biliary Self-Expanding Stent System, 135 cm Length (Catheter)- Base Part #FG1010567.

DateJun 7, 2006 (initiated Mar 22, 2006)
StatusTerminated
ReasonDesign : increasing number of customer complaints with a trend of stent deployment difficulties. The difficulties are reported during acute deployment of the stent, therefore patients in whom a stent has successfully been implanted are not affected by this action.
Root causeOther
ActionIdentify all customers who have been shipped affected product 2) Deliver Physician Medical Device Recall Notice and Reconciliation Form via Federal Express 3) Guidant field representatives contact hospitals by phone and in person to reinforce letter and recall plan4) Reconcile receipt notifications and physical product returns to expected returns.
QuantityApproximately 38,203 units (all part #s) distributed US
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033393ABSOLUTE .035 SELF-EXPANDING BILIARY STENT SYSTEM
Guidant Corp., Nov 10, 2003