ClearanceBrief
Recall Z-0802-06 ยท MMP Cover, Barrier, Protective

Smiths Medical MD, Inc.

Locator-Wand Cover that is sold in a shelf carton (unit packaging) separate to the sterile tray containing and identified as PORT-A-CATH P.A.S. PORT T2 Fluoro-Free, Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm O.D. tray. REF 21-4672-24. Contents: Portal with ULTRA-LOCK Connector, Radiopaque PolyFlow Polyurethane Catheter 5.8Fr with CATH-FINDER Sensor Wire, Locator-Wand cover, L

DateApr 27, 2006 (initiated Mar 29, 2006)
StatusTerminated
ReasonOne lot of Locator-Wand Cover sterile pouches may have an insufficient seal and product sterility could be compromised. The Locator-Wand Cover is supplied with the PORT-A-CATH II AND P.A.S. PORT systems. The packaged Wand-Cover is placed on the outside of the PORT-A-CATH II AND P.A.S. PORT tray.
Root causeOther
ActionAn Urgent Product Safety & Recall Notification, dated 03/29/06, was mailed to domestic customers and emailed to international customers. The letter described the issue, identified re-order number and lot numbers affected and gave 3 options. The options include to use only the sterile tray and discard the wand cover from the shelf box; remove and discard Locator-Wand Cover pouches and request replacement Locator-Wand Covers; or return the entire device for a credit or replacement.
Quantity181 trays
DistributionNationwide. AR, CA, DE, FL, IL, KY, LA, MA, MD, MI, MN, MS, NJ, NY, OK, PA, TN, TX, VA, WI. Germany, Italy, Netherlands, Singapore, Spain
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K942024PORT-A-CATH LOW PROFILE IMPLANTABLE VENOUS ACCESS SYSTEM
Pharmacia Deltec, Inc., Jul 26, 1994