Recall Z-0838-06 ยท KGC Tube, Gastro-Enterostomy

Bard Access Systems, Inc

22F Tri-Funnel Replacement Gastrostomy Tube

DateMay 3, 2006 (initiated Apr 6, 2006)
StatusTerminated
ReasonPackaging for some gastrostomy tubes was labeled with incorrect balloon inflation data. An underinflated balloon may cause misplacement, slippage or premature removal of the tube. If an underinflated balloon allows the tube to be placed, or slip into, the stomal tract and not the stomach, there is a risk of infusion of fluids into the peritoneum. This carries the risk of infection.
Root causeError in labeling
ActionConsignees were notified by letter on 04/06/2006.
Quantity236 units
DistributionNationwide, including 4 VA facilities. No military consignees. Foreign distribution to Canada, Australia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K910492BARD REPLACEMENT GASTROSTOMY TUBE
C.R. Bard, Inc., Jul 24, 1991