Recall Z-0821-06 ยท HGM System, Monitoring, Perinatal

LMS Medical Systems Ltd

CALM software, version 02.09.01 This software is intended for use as a central monitoring system with signal analysis and remote repeaters in the perinatal clinical environment. It interfaces with standard fetal and maternal monitors and provides fetal surveillance, monitoring of labor progress, charting, reporting and data analysis, display and archiving functions.

DateApr 27, 2006 (initiated Jan 11, 2005)
StatusTerminated
ReasonA deficiency in the software code can, in certain circumstances, cause data for the wrong patient to be displayed in the Tracing Review screen when a user is acknowledging an Alert
Root causeOther
ActionConsignees were first notified by telephone on 01/11/2005. A brief document was sent via email or fax on/about 01/13/2005.
Quantity8 units
DistributionAL, IL, MD, NH, OH and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.