ClearanceBrief
Recall Z-0874-06 ยท DHA System, Test, Human Chorionic Gonadotropin

Dade Behring , Inc.

Stratus CS STAT Fluorometric Analyzer BHCG TestPak (CBHCG) in vitro diagnostic.

DateMay 19, 2006 (initiated Apr 3, 2006)
StatusTerminated
ReasonFoil sealing of theTestPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.
Root causeOther
ActionCustomers were provided clear written instructions (04/06/2006) to discontinue use and discard remaining inventory of products over 2 months old. In addition, each customer is being contacted by phone as an effectiveness verification. Recalling firm is also instructing Stratus CS customers to process all samples and dilutions in duplicate.
Quantity126 cartons of 100 TestPaks
DistributionNationwide to hospitals and government accounts in addition to International accounts in Canada, Saudi Arabia, Bangladesh, UAE, India, Japan, Denmark, and Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003696STRATUS CS STAT FLUOROMETRIC ANALYZER BHCG TESTPAK,TEST SYSTEM, MODEL CBHCG, BHCG CALPAK,MODEL CBHCG-C, BHCG DILPAK, MOD
Dade Behring, Inc., Jan 25, 2001