Recall Z-1079-06 ยท JAK System, X-Ray, Tomography, Computed

Hitachi Medical Systems America Inc

Hitachi Emission Computed Tomography System Hitachi CX-4 CT (including SceptreP3 PET-CT)

DateJun 3, 2006 (initiated Feb 21, 2006)
StatusTerminated
ReasonThe device has a software anomaly which causes blank images to be created during Multiplanar Reconstruction processing or results in an inherent filming error in which a reference scale is improperly sized.
Root causeOther
ActionHitachi Medical Systems America, Inc. was informed of a software-related problem with the 50mm scale displayed on filmed images. This problem was reported to the firm by their Japanese manufacturer. In cases where the image has been magnified before filming, the 50mm scale is not correctly sized to the magnified image. The firm issued a Device Correctioin Notice to their customers on 2/21/2006. The notice concerns the referenced software problem which results in the resizing of the images, as well as a second previously discovered problem which results in blank multi-planer reconstructions
Quantity16
DistributionThe 16 devices were distributed to 11 states: Texas, Florida, Louisiana, Illinois, Connecticut, Ohio, Massachusetts, Wisconsin, Idaho, California, New Jersey
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040902HITACHI PRESTO
Hitachi Medical Systems America, Inc., Apr 21, 2004
K042428SCEPTREP3 PET/CT
Hitachi Medical Systems America, Inc., Sep 17, 2004