Recall Z-0902-06 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Beckman Coulter Inc
Unicel DxC 600/800 SYNCHRON Clinical Systems; Operating Software Version 1.0 and 1.2; Part Number: A20463 and A27331
| Date | May 25, 2006 (initiated Mar 15, 2006) |
| Status | Terminated |
| Reason | It was confirmed that the ORDAC (Over Range Detection And Correction) feature (auto & manual) does not perform correctly for the glucose cup chemistry (GLUCm) when sample type of 'Other' is selected w/operating software versions 1.0 & 1.2. Results for these samples could be elevated as much as twice the actual value. The ORDAC feature does function properly w/serum, plasma, urine & CSF. |
| Root cause | Other |
| Action | User Notification: A Product Corrective Action (PCA) letter will be sent the week of MAR 15, 2006 to all Synchron LX20, LX20PRO, LXi 725, and UniCel DxC 600/800 customers to inform them of the limitation for the use of fluid type 'Other' when programmed in combination with automatic or manual ORDAC for a GLUCm test. Customers are requested to implement one of three workarounds described in the PCA. IN PROCESS Stock Control: 1) A STOP SHIP (BRE-06016-1) was placed on all stock in inventory. DONE 2) Current customers of LX and DxC will be informed of the limitation for the use of fluid type 'Ot |
| Quantity | 394 (361 in US & 33 in Canada) |
| Distribution | Nationwide, also Canada, France, Germany, Switzerland, UK, Spain, Belgium, Italy, Netherlands, Sweden, Czech Republic and other countries within Europoe, So. Africa, Isreal Kuwait, united Arab Emirates, Mexico, Argentina, Japan, Australia, Hong Kong, China, Taiwan, Korea, Asia, Africa and the Middle |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K042291 | UNICEL DXC 600 AND 800 SYNCHRON SYSTEMS Beckman Coulter, Inc., Nov 12, 2004 |