Recall Z-0712-06 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA Inc

ECAT Emerge PET Scanners (distributed by Siemens Medical Solutions USA, Inc.), ECAT ART PET Scanners (distributed by Siemens Medical Solutions USA, Inc.) and Sceptre PET Scanners (distributed by Hitachi Medical Systems of America)

DateApr 4, 2006 (initiated Mar 28, 2006)
StatusTerminated
ReasonThere is a potential for inaccurate, elevated SUVs (standardized uptake values). On follow-up scans, the elevated SUV could be misinterpreted, and inappropriately impact a patient's treatment plan.
Root causeOther
ActionWritten notification was sent out beginning 03/28/2006 to end users to advise them of the problem. The notification also included specific work instructions to aid the end user until the patch is available and the upgrade could be performed by a service representative.
Quantity57 devices
DistributionKS, WA, CA, OH, FL, TX, Germany, Great Britain, Canada Switzerland, Turkey, Argentina, Austria, France, & Spain
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003241ECAT PET SCANNER
Cti Pet Systems, Inc., Dec 13, 2000
K940478ECAT PRT
Cti Pet Systems, Inc., Mar 22, 1994