Recall Z-0897-06 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

GE Healthcare

GE Healthcare Innova 3100 and Innova 3100 with Bolus Chasing offered as an option, Cardiovascular Imaging System.

DateMay 24, 2006 (initiated Mar 6, 2006)
StatusTerminated
ReasonIncorrect screws were used for mounting the detector which may lead to a possible part falling onto a patient.
Root causeOther
ActionGEHC is issuing a validated field action 3/6/2006 to correct the issues. Existing screws/bolts will be replaced using correct torque with screws containing cotter pins. Loctite liquid glue will be applied. Service manuals have been updated with instructions for using correct screws and torque. The field action is being implemented via Field Modification Instruction (FMI) 12033.
Quantity300 (191 domestic and 109 OUS)
DistributionNationwide including Puerto Rico and OUS to include: Australia, Belgium, Brazil, Bulgaria, Byelarus, Canada, China, Columbia, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Honduras, Hungary, Indonesia, Iraq, Ireland, Israel, India, Iran, Italy, Jordan, Japan, Republic o
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031637INNOVA 3100
GE Medical Systems, Aug 22, 2003