ClearanceBrief
Recall Z-0815-06 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled

Alpha Omega Services Inc

TIT0001-003, Implant Tube, Tongue/Soft Tissue, Puthawala/Syed:28cm Leader, 85cm Body, 15cm Draw-Down, Cable Insert, & Nylon Retaining Buttons

DateApr 27, 2006 (initiated Feb 21, 2006)
StatusTerminated
ReasonThere is a possibility that the diameter of the catheter does not meet specifications. As a result it may not be possible to correctly connect the catheter to a VariSource Quick Connect. This would prevent any treatment. Furthermore, since the problem would not be apparent until after the catheter has been implanted an unnecessary surgical implantation procedure would have taken place.
Root causeOther
ActionRecall Notification Letter: began February 22, 2006. Depth/Scope: 100% of the customer''s who purchased the affected products will be identified and contacted with a faxed and mailed Medical Device Recall notice and followed up with a phone call by the AOS Sales & Marketing personnel. All customers will be requested return all affected products, both used and unused.
Quantity1163
DistributionNationwide and Argentina, Brasil, Canada, Israel, Korea, , Taiwan, United Kingdom & Venezuela
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.