Recall Z-1387-06 ยท JFP Electrode, Ion Specific, Calcium

Radiometer America Inc

Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzers, ABL-800 series, manufactured from 8/2004 through 9/2005.

DateAug 10, 2006 (initiated Sep 29, 2005)
StatusTerminated
ReasonUnder certain circumstances, the ABL 700 &800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and Sodium (Na).
Root causeOther
ActionThe consignees were notified of this recall by a letter, dated 9/29/2005. The letter informs the customers of this potential problem and the availability of new updated software which will correct the problem. The letter goes on to inform the customer that a Radiometer Service Engineer wil be contacting them shortly in order to arrange for a convenient time to install the software update. Along with the letter, the customer is provided a copy of an additional page and a CD concerning the function and operation of the new software, which is to be added to the customer''s Operator''s Manual.
Quantity196 devices
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041874ABL800 FLEX
Radiometer Medical Aps, Oct 8, 2004