Recall Z-1386-06 ยท JFP Electrode, Ion Specific, Calcium

Radiometer America Inc

Blood Gas and Co-Ox2 Electrolyte & Metabolyte Analyzers, ABL-700 Series, manufactured from 8/1998 through 11/2004.

DateAug 10, 2006 (initiated Sep 29, 2005)
StatusTerminated
ReasonUnder certain circumstances, the ABL 700 & 800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and Sodium (Na).
Root causeOther
ActionThe consignees were notified of this recall by a letter, dated 9/29/2005. The letter informs the customers of this potential problem and the availability of new updated software which will correct the problem. The letter goes on to inform the customer that a Radiometer Service Engineer wil be contacting them shortly in order to arrange for a convenient time to install the software update. Along with the letter, the customer is provided a copy of an additional page and a CD concerning the function and operation of the new software, which is to be added to the customer''s Operator''s Manual.
Quantity1615 devices
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002290ABL-700 SERIES, MODEL ABL 7XX (VARIOUS MODELS)
Radiometer Medical A/S, Sep 13, 2000
K980130ABL700 SERIES BLOOD GAS , CO-OXIMETRY, ELECTROLYTE & METABOLITE SYSTEM
Radiometer America, Inc., Feb 9, 1998
K991417ABL 735 ANALYZER
Radiometer America, Inc., Nov 24, 1999
K992859ABL700 SERIES ANALYZER WITH AUTOCHECK MODULE
Radiometer America, Inc., Jan 10, 2000