Recall Z-0870-06 ยท FRN Pump, Infusion

Excelsior Medical Corp

ESP 140 Syringe Pump. Used for medication infusion. Model 140A and 140D, manufactured by Excelsior Medical Corporation, Neptune, NJ.

DateMay 17, 2006 (initiated Mar 21, 2006)
StatusTerminated
ReasonPotential for alarm failure when the 140 pump completes infusion and/or the pump encounters a mid-infusion occlusion, when used in the 140 mode with a 140cc syringe. The ESP Pump works correctly when used in 60 mode with a 60cc syringe or smaller.
Root causeOther
ActionAn 'Urgent Recall' letter was mailed certified return receipt mail on 3/22/06 to all customers per distribution list.
Quantity3911 units
DistributionNationwide and Quebec, Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K914988500 INFUSER
Excel Medical, Inc., Apr 6, 1992