Recall Z-0533-06 ยท LLZ System, Image Processing, Radiological

Toshiba American Med Sys Inc

Toshiba America Medical Systems, Ultimax System MDX-8000A equipped with the EPS-Plus/ADR-1000A/E2D Digital Spot Film Device

DateMar 29, 2006 (initiated Feb 23, 2006)
StatusTerminated
ReasonWhen the error condition occurs the x-ray exposures will terminate immediately. The digital system (EPS-Plus/ADR-1000A) will immediately stop acquiring images or will continue to acquire blank images. The remote console will lock-up upon release of the hand switch. The operator must reboot the system. This will result in the loss of the images acquired during the contrast portion of the study.
Root causeOther
ActionRecall letter to be sent to to customers of affected devices beginning February 23, 2006. Corrected software to be issued by Toshiba America Medical Systems Inc, for all affected devices.
Quantity5
DistributionNationwide to AZ, CA, FL, MT, & PA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K961307SMART SPOT