Recall Z-0710-06 ยท MKJ Automated External Defibrillators (Non-Wearable)

Medtronic Emergency Response Systems, Inc.

LIFEPAK 12 defibrillator/monitor

DateApr 4, 2006 (initiated Mar 22, 2006)
StatusTerminated
ReasonTherapy connectors may have connector sockets that do not meet design specs - short drill depth. Interruption of the connections may cause unexpected, intermittent or complete loss of pacing and defibrillation functions.
Root causeNonconforming Material/Component
ActionOn 3/22/06, the firm began sending out letters dated March 2006, and flagged URGENT MEDICAL DEVICE CORRECTION. The letters advised customers of the issue and that a service representative would visit and inspect the therapy connector sockets, and replace the therapy connector and the mating cable if short drill sockets are found.
Quantity942 devices
DistributionNationwide, and Canada, Netherlands, Germany, Australia, Guam, Hong Kong, New Zealand, New Caledonia, Argentina, Brazil, Chile, Costa Rica, Mexico, Uruguay, Venezuela. Domestic distribution includes hospitals, clinics,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010918LIFEPAK 12 DEFIBRILLATOR/MONITOR SYSTEM
Medtronic Physio-Control Corp., Aug 23, 2001