Recall Z-0809-06 ยท IZI System, X-Ray, Angiographic
Siemens Medical Solutions USA, Inc
AXIOM Artis X-ray systems running software versions VB11K and VB22G
| Date | Apr 26, 2006 (initiated Mar 10, 2006) |
| Status | Terminated |
| Reason | The recalling firm became aware of a potential problem when imaging patients on their stomach on the AXIOM Artis X-ray Systems running software versions VB11K and VB22G. In some instances, the orientation labels could be incorrect. |
| Root cause | Other |
| Action | The recalling firm has issued a Customer Safety Advisory to the affected customers per Update Instructions AX011/06/S and AX013/06/S, on 3/10/2006. This informs customers of the potential problem and provides a recommendation to avoid this issue. |
| Quantity | 287 |
| Distribution | Nationwide. The product was shipped to medical facilities in AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL,IN, KS, KY, LA, MA, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WI and WV. |