ClearanceBrief
Recall Z-0621-06 ยท HGI Electrocautery, Gynecologic (And Accessories)

Apple Medical Corporation

Fischer Cone Biopsy Excisor Model: 900-158

DateApr 6, 2006 (initiated Mar 10, 2006)
StatusTerminated
ReasonChange in product material may be cause for wire breaking, and tip burning and melting during procedure. Broken pieces of the device may need to be retrieved from patient.
Root causeOther
ActionApple Medical Corporation issued letters dated March 21, 2006 to direct accounts and distributors via US Mail. Distributors are requested to sub-recall. Product is to be returned to the firm and destroyed.
Quantity129.0 boxes (10 each)
DistributionNationwide Foreign: Australia, Austria, Taiwan, England, FInland, Germany, Iceland, Israel, Italy, Portugal, Hong Kong, Spain, The Netherlands, Chile.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K961460LWEET CONE BIOPSY INSTRUMENT
Apple Medical Corp., Nov 19, 1996