ClearanceBrief
Recall Z-0660-06 ยท FPA Set, Administration, Intravascular

Smiths Medical MD, Inc.

Large-Volume IV Administration Sets for use with Deltec 3000 and Deltec Micro 3100 Large Volume Infusion Pumps. Model Number 8C290 (Reorder No. 21-0304-01) Speciality Administration Set. Sterile, non-pyrogenic. Length 105 inch/267 cm, 25 ml priming volume, 20 drops/ml and 0 Injection sites, PVC cassette, non-PVC lined tubing, non-PVC drip chamber. Manufactured for Deltec SIMS Deltec, Inc., S

DateMar 18, 2006 (initiated Dec 8, 2005)
StatusTerminated
ReasonInaccurate infusion rate-administration sets which are used with selected infusion pumps may have a defect which does not allow the pump valve to close completely.
Root causeOther
ActionAffected customers were notified of the recall and asked to return or destroy any unused product.
Quantity7 boxes of 30 Admin. Sets (210 sets in total)
DistributionSouth Carolina, Australia, Canada and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K811933GUARDIAN, VOLUMETRIC CONTROL DELIVERY
Avi, Inc., Jul 20, 1981
K991599CADD ADMINISTRATION SET
Sims Deltec, Inc., Jun 15, 1999