Recall Z-0659-06 ยท FPA Set, Administration, Intravascular

Smiths Medical MD, Inc.

Large-Volume IV Administration Sets for use with Deltec 3000 and Deltec Micro 3100 Large Volume Infusion Pumps. ------Model Number RC4106 (Reorder No. 21-0402-25) Primary Administration Set. Sterile, non-pyrogenic. Length 105 inch/267 cm, 25 ml priming volume, 20 drops/ml and 1 needless injection site. Manufactured for Deltec SIMS Deltec, Inc., St. Paul, MN 55112 U.S.A. sims Smiths Industries

DateMar 18, 2006 (initiated Dec 8, 2005)
StatusTerminated
ReasonInaccurate infusion rate-administration sets which are used with selected infusion pumps may have a defect which does not allow the pump valve to close completely.
Root causeOther
ActionAffected customers were notified of the recall and asked to return or destroy any unused product.
Quantity4 boxes of 40 Admin. Sets (160 sets in total)
DistributionSouth Carolina, Australia, Canada and Puerto Rico.
FDA recordOpen on fda.gov