Recall Z-0676-06 ยท JIX Calibrator, Multi-Analyte Mixture

Alfa Wassermann, Inc

Gemcal Reference Serum (magnesium calibrator). Calibrator, Multianalyte Mixture. Level 1 and Level 2 Chemistry Control.

DateMar 23, 2006 (initiated Jan 26, 2006)
StatusTerminated
ReasonInaccurate test results. The magnesium values for the A31/A41 Gemcal Serum calibrator gave control results that were biased approx. 20% high.
Root causeOther
ActionField correction letters were sent to the firm's customers via FAX on 1/14/2006-1/17/2006 and via UPS on 1/17/2006-1/19/2006. An updated letter was sent out via FAX on 1/27,30,31/2006 and via UPS on 1/30/2006 and 2/1/2006.
Quantity22,987 kits
DistributionThe kits were sent nationwide. The product was shipped to Alfa Wasserman's sister firm in Wergen, Netherlands. There are several US Navy accounts.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K930104ACE CLINICAL CHEMISTRY SYSTEM
Schiapparelli Biosystems, Inc., Aug 23, 1993