Recall Z-0766-06 ยท FKP System, Dialysate Delivery, Single Patient

Baxter Healthcare Renal Div

System 1000 family of Hemodialysis Instruments, including the System 1000, TINA, AURORA and Altratouch; Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A.

DateApr 18, 2006 (initiated Feb 24, 2006)
StatusTerminated
ReasonThe older style 'Through Hardware' Dialysate Monitoring Manifolds and 'Noryl' Ultra Filtration Removal Regulators may become cracked due to mechanical and thermal stress. If either of these parts become cracked during patient treatment, there is the potential for an ultrafiltration error to occur, which can result in excessive removal of fluid from the patient.
Root causeOther
ActionBaxter sent Urgent Device Correction letters dated 02/24/06 to all System 1000, TINA, Altratouch and Aurora Hemodialysis Instrument customers via first class mail on the same date, to the attention of the Hemodialysis Administrator. The letters informed the accounts that the potential for the older style 'Through Hardware' Dialysate Monitoring Manifolds and 'Noryl' Ultra Filtration Removal Regulators to become cracked due to mechanical and thermal stress. Color photographs of the old style and new style components were included in the letter. The accounts were informed that if the cracking o
Quantity14,917 units
DistributionNationwide including Puerto Rico, and internationally to Algeria, Austria, Bahrain, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Ireland, Israel, Italy, Jordan, Lebanon, Morocco, the Netherlands, Nigeria, Oman, Poland, Romania, Russia, Slovakia, South Africa,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K910215DRAKE WILLOCK SYSTEM 1000 DIALYSATE DELIVERY SYST
Cd Medical, Inc., Apr 9, 1991