ClearanceBrief
Recall Z-0684-06 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Bayer Healthcare, LLC (Diagnostics Division)

Brand Name: RapidLab 1200 System, Model 1260, Part Number 05062460; All RapidLab 1200 Systems with Software Versions 1.0, 1.01, or 1.02. --- Common Name: Blood gases, electrolyte and blood pH test system.

DateMar 29, 2006 (initiated Jan 20, 2006)
StatusTerminated
ReasonWhen stored data is recalled from memory, the results do not agree with the original results. Software anomaly: a calculation error occurs when using the correlation adjustment feature on the RapidLab 1200 Systems with software versions 1.0, 1.01, or 1.02. The correlation feature applies user-defined slope and offset parameters each time stored test results are recalled to the display.
Root causeOther
ActionSupport Bulletins and Customer Bulletins were e-mailed to all Bayer HealthCare LLC Branches on 1/20/06 for distribution to the end user facilities (medical facilities).
DistributionUSA, Australia, Austria, Belgium, Denmark, Finland, France, Germany, Greece, Hong Kong, Italy, Japan, Korea, Malaysia, The Netherlands, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031560BAYER RAPIDLAB 1200 SYSTEM
Bayer Healthcare, LLC, Dec 17, 2003