ClearanceBrief
Recall Z-0606-06 ยท MVC System, Test, Her-2/Neu, Ihc

Biogenex Laboratories

Test kit used in the assessment of breast cancer patients for whom Herceptin therapy is being considered. BioGenex brand InSite HER-2/neu IHC Kit, Manual; Model Numbers: RD134-60K; Lot Numbers: RD1340205, RD1340205X, RD1340205Y

DateMar 15, 2006 (initiated Feb 9, 2006)
StatusTerminated
ReasonProducts produce granular staining in immunostained tissue sections, which may lead to misdiagnosis due to false positive result.
Root causeOther
ActionOn 2/9/06, the firm issued letters via UPS to all of its consignees, informing them of the affected products and providing instructions on the recall.
Quantity14 units
DistributionNationwide. Italy, and United Kingdom. The product was released for distribution to 21 consignees nationwide to CA, FL, GA, IA, IL, NC, NJ, NY, TX, VT, WA, and WI. One VA account: VA Medical Center, Gwen Levy, Bld. 500, Room 1238, 11301 Wilshire Blvd., Los Angeles, CA 90073. 2 overseas consignees
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.