Recall Z-0583-06 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc

UniCel Dxl 800 Access Immunoassay System Software Versions 2.2.1 and earlier

DateMar 3, 2006 (initiated Feb 6, 2006)
StatusTerminated
ReasonThe UniCel Dxl 800 Access Immunoassay may aspirate a sample from an incorrect rack and generate a result, which is not flagged (under a specific set of circumstances)
Root causeSoftware design
ActionA Product Corrective Action letter will be mailed the week of Feb 06 2006 to all UniCel 800 Dxl customers informing them of a rare condition on the UniCel Dxl 800 Access Immunoassay Systems. Under a specific set of circumstances, the UniCel Dxl 800 may aspirate a sample from an incorrect rack and generate a result, which is not flagged. (Letter is sent by US mail)
Quantity280 total (263 in US & 17 in Canada)
DistributionNationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023764UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM, MODEL 973100
Beckman Coulter, Inc., Jan 28, 2003