Recall Z-0612-06 ยท FRN Pump, Infusion

Baxter Healthcare Corp.

Baxter Auto Syringe AS50 Infusion Pump, product code 1M8550; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A.

DateMar 16, 2006 (initiated Feb 7, 2006)
StatusTerminated
ReasonA non-conforming gasket was used in manufacturing of the specific pumps, causing the pumps to be more susceptible to electro-static discharge, which could increase the potential for the pump to fail during use and alarm.
Root causeOther
ActionBaxter sent Urgent Device Correction letters dated 2/7/06 to the direct accounts on the same date via first class mail. The letters informed the accounts of the potential for the affected pumps to be susceptible to electro-static discharge, which could increase the potential to fail during use, causing the pump to stop infusing and issue an audible and visual alarm. The accounts were asked to immediately take the pump out of service, contact Baxter at 1-800-843-7867 to arrange for the return of the pumps in order to replace the non-conforming gasket, and complete and return the enclosed reply
Quantity282 pumps
DistributionNationwide. Arizona, California, Florida, Georgia, Hawaii, Iowa, Indiana, Kentucky, Louisiana, Michigan, Mississippi, North Carolina, Nebraska, New Mexico, New York, Ohio, Oregon, Pennsylvania, Texas, Virginia, Washington, West Virginia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K945942AUTO SYRINGE(R) AS50 INFUSION PUMP
Baxter Healthcare Corp, Jul 10, 1995