Recall Z-0661-06 ยท DQA Oximeter

Masimo Corporation

Rainbow DC-I-DC Pulse CO-Oximeter Sensor used with the Masimo SET Rad 57 Pulse CoOximeter

DateMar 18, 2006 (initiated Feb 2, 2006)
StatusTerminated
ReasonInternal testing revealed that some of the sensors might not meet the accuracy specifications of : 2% for SpO2 and of: 3% for SpCO, due to the tolerance of one of the emitter LED's used in the calculation ofSpO2 and SpCO. The SpO2 reading may have a negative bias of up to 1 % and the SpCO reading may have a negative bias of up to 4%.
Root causeOther
ActionMasimo notified all customers by letter sent via Federal Express who were shipped Rainbow DC- I-DC sensors to return them to Masimo for repair or replacement. A Tracking/Verification form will be mailed with the letter to be faxed back to Masimo. Masimo will repair or replace the returned units and return them back to the customer.
Quantity321
DistributionNationwide and Belgium, Canada, France, Germany, Italy, South Korea, new Zealand, Slovenia, Spain, Sweden, Switzerland, and United Arab Emirates
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042536MASIMO SET RAD 57 PULSE CO-OXIMETER
Masimo Corporation, Jan 19, 2005